RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778
Recall
- Recall Number
- Z-2432-2020
- Event Number
- 85750
- Firm
- Siemens Healthcare Diagnostics Inc
- FEI Number
- 3002637618
- Product Code
- CHL
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- May 27, 2020
- Terminated
- January 12, 2021
- Address
- 2 Edgewater Dr, Norwood, MA, 02062-4637
Description
RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778
The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.
On May 27, 2020, the firm notified customers via Urgent Field Safety Notice letters. Product was distributed to foreign customers only. Customers were instructed to complete and return the included customer response form. Customers whose systems are not configured to use code 39 with check-digit need take no further action. However, for customers whose systems are configured to use code 39 with check-digit, the customer notice provided two workaround options: 1) Use the integrated on-board barcode scanner, or 2) Re-configure the handheld barcode scanner following the instructions in the letter. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative
Foreign distribution only.