FDA Recall Terminated

Vectris(TM) Sure Scan(R) MRI Lead Kit for Spinal Cord Stimulation, Models: (a) 977A160 (b) 977A175 (c) 977A190 (d) 977A260 (e) 977A275 (f) 977A290 Product Usage: The Medtronic Vectris lead kits contain the implantable spinal cord stimulation lead, as well as additional components used for lead implant as part of a Neurostimulation system for pain therapy. A Medtronic implantable Neurostimulation system is indicated for spinal cord stimulation (SCS) systems as an aid in the management of chronic, intractable pain of the trunk and/or limbs-including unilateral or bilateral pain associated with the following conditions: Failed Back Syndrome (FBS) or low back syndrome or failed back; Radicular pain syndrome or radiculopathies resulting in pain secondary to FBS or herniated disk; Post laminectomy pain; Multiple back operations; Unsuccessful disk surgery; Degenerative Disk Disease (DDD)/herniated disk pain refractory to conservative and surgical interventions; Peripheral causalgia; Epidural fibrosis; Arachnoiditis or lumbar adhesive arachnoiditis; Complex Regional Pain Syndrome (CRPS), Reflex Sympathetic Dystrophy (RSD), or causalgia.

Recall: Z-2432-2018 · Initiated May 25, 2018

Recall

Recall Number
Z-2432-2018
Event Number
80256
Firm
Medtronic Neuromodulation
FEI Number
2182207
Product Code
LGW
Status
Terminated
Root Cause
Process control
Initiated
May 25, 2018
Posted
July 13, 2018
Terminated
December 11, 2020
Address
7000 Central Ave Ne, Minneapolis, MN, 55432-3568

Description

Vectris(TM) Sure Scan(R) MRI Lead Kit for Spinal Cord Stimulation, Models: (a) 977A160 (b) 977A175 (c) 977A190 (d) 977A260 (e) 977A275 (f) 977A290 Product Usage: The Medtronic Vectris lead kits contain the implantable spinal cord stimulation lead, as well as additional components used for lead implant as part of a Neurostimulation system for pain therapy. A Medtronic implantable Neurostimulation system is indicated for spinal cord stimulation (SCS) systems as an aid in the management of chronic, intractable pain of the trunk and/or limbs-including unilateral or bilateral pain associated with the following conditions: Failed Back Syndrome (FBS) or low back syndrome or failed back; Radicular pain syndrome or radiculopathies resulting in pain secondary to FBS or herniated disk; Post laminectomy pain; Multiple back operations; Unsuccessful disk surgery; Degenerative Disk Disease (DDD)/herniated disk pain refractory to conservative and surgical interventions; Peripheral causalgia; Epidural fibrosis; Arachnoiditis or lumbar adhesive arachnoiditis; Complex Regional Pain Syndrome (CRPS), Reflex Sympathetic Dystrophy (RSD), or causalgia.

Reason

This voluntary recall is being conducted due to the curved tip introducer needle which is included in the kit, to have a potential manufacturing defect. This could result in difficulty advancing or withdrawing the Vectris lead through the curved tip introducer needle.

Action

Beginning 25 May 2018, Medtronic sent a letter via mail to affected customers. The notification included the customer letter, confirmation form (reply form), and a list of affected product serial numbers. The letter informed customers of the recall, and provided instructions on the identification and return of affected products. The firm expanded the recall, notifying their consignees by letter on 07/13/2018. Customers with device related questions can contact Technical Services at 1-800-707-0933 weekdays 7am-6pm CT.

Distribution

Worldwide

Quantity

8182 units