FDA Recall
Terminated
ADVIA Chemistry Calibrator
Recall: Z-2430-2020
·
Initiated June 5, 2020
Recall
- Recall Number
- Z-2430-2020
- Event Number
- 85844
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2432235
- Product Code
- NDW
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- June 5, 2020
- Terminated
- November 18, 2021
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005
Description
ADVIA Chemistry Calibrator
Reason
Calibration errors have been observed with the Total and Direct Bilirubin assays when using Chemistry Calibrator lot 534177 for ADVIA Chemistry and lot 534179 for Atellica CH.
Action
Siemens Healthcare Diagnostics sent a letter notification sent to customers dated June 10, 2020. As one of the short-term alternatives for customers, where regulations permit, Siemens has evaluated the use of the lyophilized RANDOX Calibration Serum Level 3 Lot 1024UE. (CAL 3) for generating DBIL_2/DBil_2 and TBIL_2/TBil_2 results until a new lot of unaffected Siemens Chemistry Calibrator is available.
Distribution
US Nationwide Distribution
Quantity
1872 units