FDA Recall Open, Classified

Medline convenience kit: HERNIA REPAIR TOTE DYNJ908337D

Recall: Z-2427-2026 · Initiated April 27, 2026

Recall

Recall Number
Z-2427-2026
Event Number
98951
FEI Number
1417592
Product Code
PAD
Status
Open, Classified
Root Cause
Process control
Initiated
April 27, 2026
Posted
June 12, 2026
Address
3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753

Description

Medline convenience kit: HERNIA REPAIR TOTE DYNJ908337D

Reason

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Action

Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.

Distribution

US Nationwide distribution. OUS distribution pending.

Quantity

49,654 kits total