FDA Recall Terminated

12/14 TAPER HIGH OFFSET NECK USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS ONLY CO-CR, REF 71352112, Smith & Nephew, Inc. Memphis, TN 38116 USA. USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS.

Recall: Z-2427-2012 · Initiated August 22, 2012

Recall

Recall Number
Z-2427-2012
Event Number
63089
Firm
Smith & Nephew Inc
FEI Number
1020279
Product Code
MEH
Status
Terminated
Root Cause
Employee error
Initiated
August 22, 2012
Posted
September 20, 2012
Terminated
February 27, 2013
Address
1450 E. Brooks Rd, Memphis, TN, 38116

Description

12/14 TAPER HIGH OFFSET NECK USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS ONLY CO-CR, REF 71352112, Smith & Nephew, Inc. Memphis, TN 38116 USA. USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS.

Reason

Modular necks were manufactured out of specification by a vendor; the taper provide is undersized.

Action

Smith & Nephew sent an "URGENT:CLASS II RECALL-HIGH OFFSET MODULAR NECK R1213 letter dated August 22, 2012 to all affected customers. The letter identified the product, problem, and the actions to be taken by the customers. Customers were instructed to cease use and distribution of the product and quarantine for return. An Inventory Return Certification Form was included for customers to complete and return via fax to 901-566-7975.

Distribution

Worldwide Distribution-USA (nationwide) including the states of CA, FL, GA, IL, IN, MA, NC, NJ, NY, OH, OK, PA, TN, TX, and WA, and the country of Canada.

Quantity

70 units