FDA Recall Open, Classified

RayStation standalone software treatment planning system (RayStation 8A, 8B, 9A, and 9B, including all service packs) model no. 8.0, 8.1, 9.0, 9.1, 9.2

Recall: Z-2426-2020 · Initiated May 26, 2020

Recall

Recall Number
Z-2426-2020
Event Number
85796
Firm
RAYSEARCH LABORATORIES AB Sveavagen 9 Stockholm Sweden
FEI Number
3007774465
Product Code
MUJ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
May 26, 2020

Description

RayStation standalone software treatment planning system (RayStation 8A, 8B, 9A, and 9B, including all service packs) model no. 8.0, 8.1, 9.0, 9.1, 9.2

Reason

It is not obvious that the manual import of a treatment record into RayCare PACS does not automatically import the same data into RayTreat. In addition, there is an issue found with the display of planned and recorded Gantry Angle in the delivered beam list in RayTreat.

Action

On May 26, 2020, the firm notified affected customers of the product issue via emailed Field Safety Notices. Customers were made aware of the following issues: 1) When only a single beam record is received for a specific beam, the delivered Gantry angle is hidden and only the planned Gantry angle is shown, and 2) For users of RayCare 3B and RayStation RayTreat 9B (including all service packs), if a manual import is necessary, the treatment record must be imported manually into both RayStation and PACS. ACTIONS TO BE TAKEN BY THE USER " Be aware that the Gantry angle column in the first row for each beam in the Beam delivery results always shows the planned Gantry angle and does not indicate which Gantry angle was used when delivering a beam. " If treatment records are not successfully sent from the delivery machine to RayTreat: o always manually record all the beams that were delivered before signing off the session. o import all treatment records into both RayStation and RayPACS. " Take extra precautions when delivering a continuation session based on a manually recorded or offline imported treatment records to ensure that the correct amount of radiation is delivered during the continuation. " Educate treatment staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. The firm plans to resolve these issues in the next version of RayStation, which the firm estimates will be available in June 2020. If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use.

Distribution

Distributed to one consignee in Nashville. Foreign distribution to one consignee in Belgium.

Quantity

12