FDA Recall Completed

Ingenia Ambition X- Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Models: 781356

Recall: Z-2422-2021 · Initiated August 11, 2021

Recall

Recall Number
Z-2422-2021
Event Number
88491
Firm
Philips North America Llc
FEI Number
3016618143
Product Code
LNH
Status
Completed
Root Cause
Device Design
Initiated
August 11, 2021
Address
222 Jacobs St, Cambridge, MA, 02141-2289

Description

Ingenia Ambition X- Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Models: 781356

Reason

When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment

Action

Philips notified U.S. customers an Urgent Medical Device Correction letter on 8/11/21. Letter states reason for recall, health risk and action to take: Philips advises that customers continue to utilize the primary motorized function for the MR system since the identified issue with the HDU does not occur in the motorized mode. If you utilize the manual function and it is not functioning properly, then you must perform table movement using the motorized function. A Philips Field Service Engineer (FSE) will perform a HDU during the next scheduled preventive maintenance visit. If the issue is detected in the HDU, the Philips FSE will replace the necessary mechanical component(s) to resolve the issue. If you need additional information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377) and reference FCO78100527.

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Albania, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan Palestine, State of Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam.

Quantity

147 units