FDA Recall
Terminated
Getinge 86 Series Washer/Disinfector with G1 control system software, Model 8666 and Model 8668. For use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.
Recall: Z-2419-2015
·
Initiated June 26, 2015
Recall
- Recall Number
- Z-2419-2015
- Event Number
- 71777
- Firm
- Getinge Disinfection Ab Ljungadalsgatan
- FEI Number
- 3000231474
- Product Code
- MEC
- Status
- Terminated
- Root Cause
- Software Manufacturing/Software Deployment
- Initiated
- June 26, 2015
- Posted
- August 25, 2015
- Terminated
- May 23, 2018
- Address
- 11 Vaxjo Sweden
Description
Getinge 86 Series Washer/Disinfector with G1 control system software, Model 8666 and Model 8668. For use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.
Reason
Getinge Disinfection AB has initiated a Field Correction due to the potential for unauthenticated remote access to the product over the network.
Action
Getinge Disinfection AB has initiated a Field Correction due to unauthenticated remote access to the product over the network.
Distribution
Distributed in CA, CT, GA, IA, ID, MD, MI, MN, NY, PA, TX, and WA.
Quantity
20 devices