Medline convenience kits: LOOP PACK DYNJ61460C OPEN HEART CDS840150Y PACEMAKER DYNJ56297C PACEMAKER PK DYNJT3454 PACEMAKER/ICD PACK DYNJ33822L PAD OPEN HEART DYNJ900962O PEDIATRIC OPEN HEART DYNJ54238L DYNJT6766 STAZ BASIC VASCULAR DYNJ909794C VASCULAR ACCESS KIT CVI5323A VASCULAR CDS CDS983040M
Recall
- Recall Number
- Z-2418-2026
- Event Number
- 98951
- FEI Number
- 1417592
- Product Code
- OEZ
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- April 27, 2026
- Posted
- June 12, 2026
- Address
- 3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753
Description
Medline convenience kits: LOOP PACK DYNJ61460C OPEN HEART CDS840150Y PACEMAKER DYNJ56297C PACEMAKER PK DYNJT3454 PACEMAKER/ICD PACK DYNJ33822L PAD OPEN HEART DYNJ900962O PEDIATRIC OPEN HEART DYNJ54238L DYNJT6766 STAZ BASIC VASCULAR DYNJ909794C VASCULAR ACCESS KIT CVI5323A VASCULAR CDS CDS983040M
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
US Nationwide distribution. OUS distribution pending.
49,654 kits total