FDA Recall Open, Classified

Arial Call Station, Model Numbers CSK200 and CSK200MR, is a wireless, battery-operated call device installed inside resident living areas to allow immediate staff response.

Recall: Z-2417-2024 · Initiated June 20, 2024

Recall

Recall Number
Z-2417-2024
Event Number
94878
Firm
Securitas Healthcare LLC
FEI Number
1929691
Product Code
IQA
Status
Open, Classified
Root Cause
Software Manufacturing/Software Deployment
Initiated
June 20, 2024
Posted
July 23, 2024
Address
4600 Vine St, Lincoln, NE, 68503-2823

Description

Arial Call Station, Model Numbers CSK200 and CSK200MR, is a wireless, battery-operated call device installed inside resident living areas to allow immediate staff response.

Reason

There is a risk that alarms will not be received by the Arial Wireless Emergency Call System and premature depletion of the battery.

Action

Securitas Healthcare issued a recall notice on 06/19/2024 to its consignees via USPS certified mail. The notice explained the problem, potential hazard, and requested the following: "Customer Required Actions: 1. Reduce the Missing Device Timeout to 125 Minutes Hours for all Call Stations. Clients should reduce the Missing Device Timeout to 125 minutes so that a Missing Device Alert will be triggered if a Call Station fails to send a supervision message to the Arial Emergency and Nurse Call System within 125 minutes from the last supervision message. Instructions for reducing the Missing Device Timeout to 125 minutes are attached hereto as Appendix 1. 2. Closely Monitor Missing Device Alerts within the Arial Web Application and Immediately Verify Status of Missing Devices and Cease Using any Call Stations That Do Not Reconnect. Clients should closely monitor the "Missing Device Alerts within the Arial Web Application and immediately check the status of Call Stations that trigger a Missing Device Alert and replace the Call Station battery if it is low or fully depleted. Instructions for replacing the Call Station battery are attached hereto as Appendix 2. Instructions for reconnecting missing devices and confirming that the Call Stations are properly transmitting information to the Arial Emergency and Nurse Call System are attached hereto as Appendix 3. Clients should immediately cease using any Call Station that will not reconnect to the Arial Emergency and Nurse Call System. 3. Closely Monitor the Battery Status of all Call Stations in the Arial Web Client and Replace Battery if it is Low . Clients should closely monitor the Call Station Battery Status in the Arial Web Client and immediately replace the battery if it low as described in the documentation. 4. Use Supplementary Communication Devices or Procedures for High Risk Residents. Clients should enact backup procedures, including increased room checks, or providing alternate emergency

Distribution

US: NE, MN, WA, MD, TX, ON, WI, OR, MS, CA, and Canada

Quantity

1268 units