FDA Recall Open, Classified

SlideView VS200 slide viewer,

Recall: Z-2416-2024 · Initiated July 22, 2024

Recall

Recall Number
Z-2416-2024
Event Number
95028
Firm
EVIDENT SCIENTIFIC INC
FEI Number
3023822598
Product Code
RIB
Status
Open, Classified
Root Cause
Radiation Control for Health and Safety Act
Initiated
July 22, 2024
Posted
July 26, 2024
Address
48 Woerd Ave, Waltham, MA, 02453-3826

Description

SlideView VS200 slide viewer,

Reason

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

Distribution

US