FDA Recall Open, Classified

Spectrum IQ Infusion Pump, Product Code 3570009

Recall: Z-2415-2025 · Initiated July 14, 2025

Recall

Recall Number
Z-2415-2025
Event Number
97335
Firm
Baxter Healthcare Corporation
FEI Number
1417572
Product Code
FRN
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
July 14, 2025
Posted
August 22, 2025
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

Spectrum IQ Infusion Pump, Product Code 3570009

Reason

Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.

Action

Customers were contacted via telephone beginning 7/14/25. The included information about the product, the issue and the action to be taken by the customers. Baxter Healthcare is requesting the return of the impacted product to perform all tests required and confirm that the product perform as expected. If you have any questions, phone (Toll Free): 800-422-9837 or 847-948-4770; hours of operation: Monday through Friday from 8:00 am to 5:00 pm CST

Distribution

US distribution to states of: AZ, FL

Quantity

6 units