Spectrum IQ Infusion Pump, Product Code 3570009
Recall
- Recall Number
- Z-2415-2025
- Event Number
- 97335
- Firm
- Baxter Healthcare Corporation
- FEI Number
- 1417572
- Product Code
- FRN
- Status
- Open, Classified
- Root Cause
- Nonconforming Material/Component
- Initiated
- July 14, 2025
- Posted
- August 22, 2025
- Address
- 1 Baxter Pkwy, Deerfield, IL, 60015-4625
Description
Spectrum IQ Infusion Pump, Product Code 3570009
Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.
Customers were contacted via telephone beginning 7/14/25. The included information about the product, the issue and the action to be taken by the customers. Baxter Healthcare is requesting the return of the impacted product to perform all tests required and confirm that the product perform as expected. If you have any questions, phone (Toll Free): 800-422-9837 or 847-948-4770; hours of operation: Monday through Friday from 8:00 am to 5:00 pm CST
US distribution to states of: AZ, FL
6 units