FDA Recall Terminated

DuraspanTM Dialysis Catheters, Ultra and Long Term, 15.5 Fr, Sterile EO, Manufactured for r4 Vascular, Inc . 7550 Meridan Circle North, Suite 150, Maple Grove, MN 55369. Description Model # Duraspan Kit - 24cm 62152420 Duraspan Kit - 28cm 62152820 Duraspan Kit - 32cm 62153220 Duraspan Exchange Kit - 32cm 62153228 Duraspan Kit - 36cm 62153620 Duraspan Exchange Kit - 36cm 62153628 Duraspan Kit - 40cm 62154020 Duraspan Exchange Kit - 40cm 62154028 Duraspan Kit - 55cm 62155520 Duraspan Ultra Kit - 24cm 621524E0 Duraspan Ultra Kit - 28cm 621528E0 Duraspan Ultra Kit - 36cm 621536E0 Duraspan Ultra Kit - 40cm 621540E0 Duraspan Ultra Kit - 55cm 621555E0 The r4 DuraspanTM and the DuraspanTM Ultra long-term dialysis catheter are indicated for attaining short and long term (>30 days) vascular access for hemodialysis, hemoperfusion or apheresis therapy. Access is attained via the internal jugular vein, external jugular vein, subclavian vein or femoral vein. Catheters 40 cm and longer are for femoral vein insertion. The ability of the Biomimetic Coating to reduce platelet adhesion and thrombus accumulation is supported exclusively by in-vitro and animal testing.

Recall: Z-2412-2011 · Initiated May 3, 2011

Recall

Recall Number
Z-2412-2011
Event Number
58735
Firm
R4 Vascular Inc
FEI Number
3006242715
Product Code
MSD
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
May 3, 2011
Posted
June 6, 2011
Terminated
July 27, 2012
Address
7550 Meridian Cir N, Ste 150, Maple Grove, MN, 55369-4932

Description

DuraspanTM Dialysis Catheters, Ultra and Long Term, 15.5 Fr, Sterile EO, Manufactured for r4 Vascular, Inc . 7550 Meridan Circle North, Suite 150, Maple Grove, MN 55369. Description Model # Duraspan Kit - 24cm 62152420 Duraspan Kit - 28cm 62152820 Duraspan Kit - 32cm 62153220 Duraspan Exchange Kit - 32cm 62153228 Duraspan Kit - 36cm 62153620 Duraspan Exchange Kit - 36cm 62153628 Duraspan Kit - 40cm 62154020 Duraspan Exchange Kit - 40cm 62154028 Duraspan Kit - 55cm 62155520 Duraspan Ultra Kit - 24cm 621524E0 Duraspan Ultra Kit - 28cm 621528E0 Duraspan Ultra Kit - 36cm 621536E0 Duraspan Ultra Kit - 40cm 621540E0 Duraspan Ultra Kit - 55cm 621555E0 The r4 DuraspanTM and the DuraspanTM Ultra long-term dialysis catheter are indicated for attaining short and long term (>30 days) vascular access for hemodialysis, hemoperfusion or apheresis therapy. Access is attained via the internal jugular vein, external jugular vein, subclavian vein or femoral vein. Catheters 40 cm and longer are for femoral vein insertion. The ability of the Biomimetic Coating to reduce platelet adhesion and thrombus accumulation is supported exclusively by in-vitro and animal testing.

Reason

r4 Vascular, Inc. is conducting a voluntary recall of specific lots of Duraspan Dialysis Catheters. Duraspan catheters are 15.5 Fr dual lumen chronic hemodialysis catheters. Over a six month period, seven patients at five hospitals experienced separation of the luer connector from the extension tube on the Duraspan catheter. Testing during failure analysis disclosed that these lots and other lot

Action

R4 Vascular Inc sent an Urgent Medical Device Recall letter dated May 3, 2011, to all affected customers. The letter described the problem and the product being recalled. Consignees were instructed to stop distributing the product and to quarantine the recalled lots and if they had already distributed the affected product to seek its return from their customers. A template letter to physicians and healthcare facilities was enclosed for Distributors/Retailers to send or hand deliver to their customers. For questions regarding this recall call 1-704-905-2105.

Distribution

Nationwide Distribution including CA, CO, CT, FL, GA, IL, LA, MD, MA, MN, MS, MO, NJ, NC, OH, PA, SC, TX, and WA.

Quantity

2489 kits