FDA Recall Terminated

Eleganza 3 Hospital Bed Intended use: AC Powered adjustable Hospital Bed

Recall: Z-2411-2011 · Initiated May 2, 2011

Recall

Recall Number
Z-2411-2011
Event Number
58785
Firm
Linet Americas Inc
FEI Number
3008007426
Product Code
FNK
Status
Terminated
Root Cause
Device Design
Initiated
May 2, 2011
Posted
June 3, 2011
Terminated
July 19, 2012
Address
10430 Harris Oak Blvd, Ste A, Charlotte, NC, 28269-7511

Description

Eleganza 3 Hospital Bed Intended use: AC Powered adjustable Hospital Bed

Reason

Linet has been made aware that there has been breakage of the Mobi-lift handle on a few beds which may cause failure during patient assist use.

Action

The firm, Linet America, notified US customers only via phone on May 5, 2011. A follow up "service advisory information" letter dated May 11, 2011 was sent to all customers. The letter described the product, problem and action to be taken. The customers were instructed to remind/retrain staff to ensure that the Mobi-lift handles are always in the stowed position during patient transport and after use, and complete and return the attached Inspection procedure for each affected Eleganza 3 bed via mail to: Linet Americas, 10420-R Harris Oaks Blvd., Charlotte, NC 28269 or fax to: 704-248-5655. Linet Americas will be receiving a supply of new Mobi-lift design and will begin the process of executing service calls to replace the Morbi-lift handles with the new design. If you have any questions, contact the Director of Quality and Service Operations at 704-248-5650 or 980-406-6446.

Distribution

Worldwide distribution: USA (nationwide) including states of: KY, MA, NC, and NY; and country of: Canada.

Quantity

1601 units