ADVIA Centaur FSH Assay: ADVIA Centaur Calibrator B US: Calibrator B 00652707 (6 pack) or 00649625 (2 pack ) OUS:OUS: 2 pack REF: 00650577 and Ref: 00654130 (6 pack) Product Usage: ADVIA Centaur CP FSH assay: For in vitro diagnostic use in the quantitative determination of follicle-stimulating hormone (FSH) in serum using the ADVIA Centaur and ADVIA Centaur XP systems
Recall
- Recall Number
- Z-2410-2012
- Event Number
- 62897
- Firm
- Siemens Healthcare Diagnostics, Inc
- FEI Number
- 1219913
- Product Code
- CGJ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- August 7, 2012
- Posted
- September 19, 2012
- Terminated
- February 26, 2014
- Address
- 333 Coney St, East Walpole, MA, 02032-1516
Description
ADVIA Centaur FSH Assay: ADVIA Centaur Calibrator B US: Calibrator B 00652707 (6 pack) or 00649625 (2 pack ) OUS:OUS: 2 pack REF: 00650577 and Ref: 00654130 (6 pack) Product Usage: ADVIA Centaur CP FSH assay: For in vitro diagnostic use in the quantitative determination of follicle-stimulating hormone (FSH) in serum using the ADVIA Centaur and ADVIA Centaur XP systems
ADVIA Centaur Prolactin Assay and FSH, Calibrator B fails the analytical sensitivity stated in the IFU for lots below 40
Siemens Healthcare issued an Urgent Field Safety Notice letter dated August 2012 by e-mail to all affected Siemens Healthcare Diagnostics Branch offices on August 7, 2012. Customers were instructed to discontinue using the ADVIA Centaur Calibrator B lots 40 and below following their obtaining Calibrator 8 lot 41 or higher. Customers were asked to complete and fax-back the response form attached.
Worldwide Distribution - US Nationwide and the countries of: Canada, Argentina, Australia, Bangaldesh,Brazil, Chile, China, Columbia, Croatia, Ecuduar, Egypt, Hong Kong, India, Israel,Japan, Korea, Maylasia, New Zealand, Mexico, Peru, Philipines, Pakistan, Saudia Arabia, Singapore, Thailand, Taiwan, Viet Nam, South Africa and UK.
15,445 kits