FDA Recall Open, Classified

Infinity FMS Pak

Recall: Z-2408-2023 · Initiated June 28, 2023

Recall

Recall Number
Z-2408-2023
Event Number
92698
Firm
Alcon Research, LTD.
FEI Number
2523835
Product Code
HQC
Status
Open, Classified
Root Cause
Process control
Initiated
June 28, 2023
Posted
August 16, 2023
Address
714 Columbia Ave, Sinking Spring, PA, 19608-1405

Description

Infinity FMS Pak

Reason

Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

Action

Customers were sent a Voluntary Medical Device Correction Notification by mail and dated 06/29/2023. Customers are asked to follow the risk mitigation precautions detailed in the notice to avoid the instance of overtightening the Phaco Tip and generating plastic particles. Additionally, customers are asked to forward the recall notification to all departments or organizations that use recalled devices. Customers are asked to complete and return the provided Response Form by fax at 817-302-4337 or email at [email protected]. Customers who have experienced any adverse events or quality related issues are to contact alcon via email at [email protected] or by phone at 1-800-757-9780, Option 5.

Distribution

Worldwide distribution - US Nationwide and the countries of Aland Islands, Albania, Algeria, Andorran, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bermuda, Botswana, Brazil, Brunei Daruss., Bulgaria, Burma, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, French Guayana, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Isle of Man, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, Nicaragua, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Fed., Saint Martin, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad, Tobago, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, USA, USVI, Utd. Arab Emir., Uzbekistan, Vietnam, and Yemen.

Quantity

51,054 units