FDA Recall Terminated

Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT 2, Product Code IAP-0400

Recall: Z-2408-2020 · Initiated May 20, 2020

Recall

Recall Number
Z-2408-2020
Event Number
85684
Firm
Arrow International Inc
FEI Number
3015859709
Product Code
DSP
Status
Terminated
Root Cause
Component design/selection
Initiated
May 20, 2020
Posted
June 26, 2020
Terminated
January 7, 2023
Address
2400 Bernville Rd, Reading, PA, 19605-9607

Description

Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT 2, Product Code IAP-0400

Reason

A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.

Action

Urgent Medical Device Correction notification letters dated 5/20/20 were sent to customers. ACTIONS TO BE TAKEN BY FACILITIES (Immediate): Please immediately check your inventory of Arrow AutoCAT2 and Arrow AC3 Optimus IABPs, whether stored or in use, and determine if you have an IABP with a model number listed above. Please see Appendix 1 for product codes and serial numbers. If an IABP in the scope of this Medical Device Correction displays a System Error 3 or High Baseline alarm, now or at any point in the future until the Long-Term Corrective Action detailed below occurs, immediately quarantine the device and contact Teleflex at 1-855-419-8507 or [email protected] to receive support for inspection and servicing of the impacted device. We recommend against using an IABP for ground or air transport between medical facilities pending implementation of the Customer Action (Long Term). For IABP use during ground or air transport prior to implementation of the Customer Action (Long Term), it is recommended that the risks and benefits of using the IABP be assessed by the medical team treating the patient and that alternative circulatory support devices be considered. Customer Action (Subsequent): Teleflex will contact each impacted facility to schedule inspection and servicing of all IABP units that contain this internal component. This inspection and servicing will be performed onsite at facilities, for all IABP units. Facilities should adhere to the following instructions pending implementation of the Teleflex Long-Term Corrective Action described later in this letter: 1. Ensure that a backup IABP is available as instructed within the Operator Manual. If no such replacement IABP is immediately available, it is recommended that the risks and benefits of using the IABP be assessed by the medical team treating the patient and that alternative circulatory support devices be considered. 2. IABP units should be closely monitored during delivery of IAB

Distribution

Worldwide distribution.

Quantity

2123 total