Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10

Recall: Z-2406-2021 · Initiated July 26, 2021

Recall

Recall Number
Z-2406-2021
Event Number
88451
Firm
Elekta Inc
FEI Number
1937649
Product Code
MUJ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
July 26, 2021
Address
1450 Beale St, Ste 205, Saint Charles, MO, 63303

Description

Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10

Reason

If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.

Action

Elekta will send an Important Field Safety Notice 382-01-MON-019 to all affected customers beginning August 10, 2021. The notice informs users of the specific product and version numbers affected by the issue, and any workaround that can be used to avoid the issue. A copy of the Important Field Safety Notice is to be kept with the most current labeling and all personnel working with the product should be made aware of the content of the letter. Customers are asked to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta representative as soon as possible, but no later than 30 days after receipt. This issue is expected to be addressed in a future Monaco release, tentatively scheduled for the end of June 2022.

Distribution

Distribution in US - IA, NJ, NY, TN, PA, TX, and WI OUS distribution in Australia, Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Japan, Korea, Netherlands, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom