MEVION S250i, MEVION S250; Proton Radiation Treatment System
Recall
- Recall Number
- Z-2406-2020
- Event Number
- 85748
- Firm
- Mevion Medical Systems, Inc.
- FEI Number
- 3007087027
- Product Code
- LHN
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- May 15, 2020
- Terminated
- December 11, 2020
- Address
- 300 Foster St, Littleton, MA, 01460-2017
Description
MEVION S250i, MEVION S250; Proton Radiation Treatment System
Couch Correction moves sent after using a 3D CT scan are partially lost if the previous correction sent or recorded had some of the Couch Correction checkboxes turned off
Mevion issued letter (pCAR0012) on May 15, 2020 via mail or FedEx, with return receipt confirmation. Letter states reason for recall, health risk and action to take: Until a software fix is applied for this problem, the user is cautioned to carefully pay attention to the 3 indicators of Couch corrections. If they disagree with the intended moves the user may simply repeat the alignment step and the problem should clear. Required User Action When using multimodal imaging the user should always check the indicators for Couch Corrections before they are applied. This includes the Verity Couch Correction widget, the Verity dialog box the informs the user of the actions being sent to the Couch, and the In Room Monitor lower right panel that displays the moves before they are made. If these disagree with the intended moves, the user should back up a step and resend the corrections. Users must confirm positioning for every applied Couch correction that all intended moves are indicated and made.
Distribution to US in MO, NJ, OH, OK, FL, and DC, and the Netherlands
8 units