FDA Recall
Terminated
Cell Search Circulating Tumor Kit. Intended for the enumeration of circulating tumor cells (CTC) of epithelial origin in whole blood.
Recall: Z-2406-2015
·
Initiated May 15, 2015
Recall
- Recall Number
- Z-2406-2015
- Event Number
- 71673
- Firm
- Veridex, LLC
- FEI Number
- 3004619490
- Product Code
- NQI
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- May 15, 2015
- Posted
- August 18, 2015
- Terminated
- March 30, 2016
- Address
- 1001 US Route 202, Raritan, NJ, 08869-0606
Description
Cell Search Circulating Tumor Kit. Intended for the enumeration of circulating tumor cells (CTC) of epithelial origin in whole blood.
Reason
Complaints of the presence of artifacts that appears as small bead like structures in the image gallery when processing both control cells and patient samples as well as the failure of high control cells.
Action
Janssen Diagnostics, LLC sent a urgent product correction notification/customer confirmation of receipt on 5/15/2015 via Fed Ex.
Distribution
Nationwide Distribution.
Quantity
1560 kits