Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software
Recall
- Recall Number
- Z-2405-2026
- Event Number
- 98963
- Firm
- Abbott Medical
- FEI Number
- 2017865
- Product Code
- PNJ
- Status
- Open, Classified
- Root Cause
- Software Design Change
- Initiated
- May 12, 2026
- Posted
- June 11, 2026
- Address
- 15900 Valley View Ct, Sylmar, CA, 91342-3577
Description
Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software
Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.
On May 12 Abbott issued a Urgent Medical Device Recall notification via sales representative. Abbott ask consignees to take the following actions: Abbott has developed updated Merlin PCS 3650 programmer software (v28.9.1 rev 1 or higher) that resolves this behavior. Beginning with this notification, your Abbott Representative will upgrade programmer software to this software version . Abbott's service representative will proved a attestation to acknowledge receipt of the notification and software update. Please share this notification with others in your organization, as appropriate.
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. OUS: Not provided yet
11182 programs