Synchron Acetaminophen (ACTM) Reagent, Part Number 472169. ACTM Reagent when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 Systems and SYNCHRON Systems Drug Calibrator 2 set, is intended for quantitative determination of Acetaminophen concentration in human serum or plasma.
Recall
- Recall Number
- Z-2405-2008
- Event Number
- 49069
- Firm
- Beckman Coulter Inc
- FEI Number
- 2050012
- Product Code
- LDP
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- July 10, 2008
- Posted
- September 25, 2008
- Terminated
- July 11, 2012
- Address
- 200 S Kraemer Blvd, Brea, CA, 92822-6208
Description
Synchron Acetaminophen (ACTM) Reagent, Part Number 472169. ACTM Reagent when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 Systems and SYNCHRON Systems Drug Calibrator 2 set, is intended for quantitative determination of Acetaminophen concentration in human serum or plasma.
This recall was initiated after Beckman Coulter verified through testing that Acetaminophen reagents (lots #M702524 & M703276) manufactured with contaminated heparin has shown a negative bias in performance between the assay made with contaminated and uncontaminated heparin.
The initial customer notifications with fax back response forms were mailed via US Postal Service beginning on 07/21/2008. The letter asked their customers to please discontinue use and discard appropriately the affected lots of Synchron Acetaminophen Reagent (M703276 & M702524). Customers were also instructed to contact their Customer Service Representative at (800) 526-3821 in the United States or (800) 463-7828 in Canada or, if they are outside the US or Canada, please contact your local Beckman Coulter Representative to discuss replacement options. They were also instructed to complete and return the enclosed response form within 10 days. In addition, the customers were informed that the contaminated Heparin was used in the following Beckman Coulter, Inc., products: Synchron Theophylline (THE) Reagent (lot # M701460), Synchron Phenytoin (PHY) Reagent and Synchron C-Reactive Protein (CRP) Reagent (lot# M703222, M703266, M703267 & M703268), however Beckman Coulter has verified through testing that the performance of these products are not affected.
Nationwide and Canada
7764 test kits