FDA Recall Open, Classified

Azurion

Recall: Z-2403-2023 · Initiated July 19, 2023

Recall

Recall Number
Z-2403-2023
Event Number
92855
Firm
Philips Medical Systems Nederland B.V. High Tech Campus 36 Eindhoven Netherlands
FEI Number
3000976525
Product Code
OWB
Status
Open, Classified
Root Cause
Radiation Control for Health and Safety Act
Initiated
July 19, 2023
Posted
September 7, 2023

Description

Azurion

Reason

A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation

Action

Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide actions that should be taken by the customer/user in order to prevent risks to patients. Philips will contact customers to schedule a visit to inspect the foot switch and provide an addendum to the Instructions for Use (IFU) of the system detailing the appropriate cleaning and maintenance instructions. (Philips reference 2023-IGT-BST-011).

Distribution

US Nationwide Distribution

Quantity

5432 systems in total