FDA Recall Terminated

Winco Elite Care Cliner w/Swing Away Arms Standard - with Nylon Casters Oncology, Dialysis, Hospital Recovery, Home Dialysis Use, Physician Facilities.

Recall: Z-2400-2015 · Initiated April 2, 2015

Recall

Recall Number
Z-2400-2015
Event Number
71496
Firm
Winco Mfg., LLC
FEI Number
1027229
Product Code
INN
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 2, 2015
Posted
August 14, 2015
Terminated
December 14, 2016
Address
5516 SW 1st Ln, Ocala, FL, 34474-9307

Description

Winco Elite Care Cliner w/Swing Away Arms Standard - with Nylon Casters Oncology, Dialysis, Hospital Recovery, Home Dialysis Use, Physician Facilities.

Reason

Shipment of foam for chairs was received classified as Fire Retardant but was in fact standard foam and would not meet Fire Retardant standards.

Action

Winco sent a field correction letter dated April 2, 2015, to all affected customers The letter identified the product, the problem, and the action to be taken by the customer. Winco instructed customers that they will replace the recliners at no charge to the customer. Winco will provide detailed instructions for the replacement of the seats by their facility staff. Winco will extend the remaining warranty on the entire chair for a period of 6 months as compensation. If product was further distributed customers were instructed to notify those customers of the field correction. Customers were asked to complete and return the enclosed field correction response form as soon as possible. Customers with questions should contact Customer Care Monday-Friday 8:30 AM-5:00 PM EST at 800-237-3377 or email [email protected]. For questions regarding this recall call 352-854-2929.

Distribution

Nationwide Distribution including CA, FL, KS, KY, LA, MA, MD, MN, MS, NJ, NY, OK, TN, TX, and WA.

Quantity

32 chairs