FDA Recall Terminated

Stratus CS Acute Care Troponin I TestPak ( Troponin I assay)

Recall: Z-2400-2010 · Initiated April 13, 2010

Recall

Recall Number
Z-2400-2010
Event Number
55610
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
MMI
Status
Terminated
Root Cause
Process control
Initiated
April 13, 2010
Posted
September 11, 2010
Terminated
October 25, 2010
Address
500 GBC Drive PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101

Description

Stratus CS Acute Care Troponin I TestPak ( Troponin I assay)

Reason

False elevated Stratus CS CTNI results

Action

Recalling firm issued " Urgent Field Safety Notice" dated April 2010, informing users of the problem and to discuss with laboratory directors regarding the need to review previous test results, conduct patient follow up, and/or repeat testing for tests conducted on or after Nov 1, 2009.

Distribution

Nationwide Distribution

Quantity

7500 cartons