Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay
Recall
- Recall Number
- Z-2398-2026
- Event Number
- 98985
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2517506
- Product Code
- MLM
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- May 7, 2026
- Posted
- June 10, 2026
- Address
- 500 Gbc Dr, Newark, DE, 19702-2466
Description
Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay
A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.
Siemens Healthineers issued an Urgent Medical Device Correction notice to its consignees on 5/12/2026 via FedEx. The notice described the issue, its impact to QC and patient results, and requested the following: "Customer Actions . Immediately discontinue use of and discard the affected kit lot numbers. . Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. . Review your inventory of these products to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for required regulatory reporting. . Complete and return the Field Correction Effectiveness Check Form attached to this letter within (30) days. . Please retain this letter with your laboratory records and forward it to anyone who may have received or used this product." For questions, contact your Siemens Healthineers Customer Care Center at (312) 275-7795 or your local Siemens Healthineers technical support representative.
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Brazil, Croatia, Germany, Hungary, India, Italy, Japan, Korea, Lithuania, Pakistan, Poland, Saudi Arabia, Spain, South Africa, Sri Lanka, Turkey, United Arab Emirates.
2878 units