Zyston Transform Instrument Case. Model Number PCR8200-2101. Device labeled sterile.
Recall
- Recall Number
- Z-2397-2016
- Event Number
- 74687
- Firm
- Zimmer Biomet Spine, Inc
- FEI Number
- 3012447612
- Product Code
- MAX
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- June 30, 2016
- Terminated
- March 15, 2017
- Address
- 310 Interlocken Pkwy, Suite 120, Broomfield, CO, 80021-3479
Description
Zyston Transform Instrument Case. Model Number PCR8200-2101. Device labeled sterile.
Zimmer Bioment Spine, Inc. announces a voluntary field action for the Zyston Transform Instrument Case and Zyston Transform Implant Kit because the locking mechanism on the implant is not functioning as intended.
Verbal notification of the decision to recall the product and related instructions was provided on June 30, 2016. Additionally, written notification of the decision to recall the product and related instructions will be provided to each of the individuals responsible for locations that have affected inventory under their control. Clear instructions and materials necessary to support the accountability and retrieval of product from the field will be provided with the recall notification. In order to ensure that the locations were alerted of the recall in a timely fashion, supplemental email notifications will also be sent to each recall notification addressee before distributing the recall documentation via Federal Express.
Nationwide to PA, TX, TN, NV, VA, WV, and KY. No Canadian, other foreign, VA/gov/military consignees.
5