FDA Recall Terminated

Zyston Transform Instrument Case. Model Number PCR8200-2101. Device labeled sterile.

Recall: Z-2397-2016 · Initiated June 30, 2016

Recall

Recall Number
Z-2397-2016
Event Number
74687
Firm
Zimmer Biomet Spine, Inc
FEI Number
3012447612
Product Code
MAX
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
June 30, 2016
Terminated
March 15, 2017
Address
310 Interlocken Pkwy, Suite 120, Broomfield, CO, 80021-3479

Description

Zyston Transform Instrument Case. Model Number PCR8200-2101. Device labeled sterile.

Reason

Zimmer Bioment Spine, Inc. announces a voluntary field action for the Zyston Transform Instrument Case and Zyston Transform Implant Kit because the locking mechanism on the implant is not functioning as intended.

Action

Verbal notification of the decision to recall the product and related instructions was provided on June 30, 2016. Additionally, written notification of the decision to recall the product and related instructions will be provided to each of the individuals responsible for locations that have affected inventory under their control. Clear instructions and materials necessary to support the accountability and retrieval of product from the field will be provided with the recall notification. In order to ensure that the locations were alerted of the recall in a timely fashion, supplemental email notifications will also be sent to each recall notification addressee before distributing the recall documentation via Federal Express.

Distribution

Nationwide to PA, TX, TN, NV, VA, WV, and KY. No Canadian, other foreign, VA/gov/military consignees.

Quantity

5