FDA Recall Terminated

BD FACS 7-color Setup beads BD Biosciences, San Jose, CA 95131. BD FACS 7 color Setup Beads are for in vitro diagnostic use on BD FACSCanto or BD FACSCanto II systems. The beads are run with BD FACSCanto clinical software. This system provides application specific setup, including automated compensation and cytometer quality control (QC) with minimal user intervention.

Recall: Z-2397-2012 · Initiated June 29, 2012

Recall

Recall Number
Z-2397-2012
Event Number
62494
Firm
BD Biosciences, Systems & Reagents
FEI Number
2916837
Product Code
GKZ
Status
Terminated
Root Cause
Employee error
Initiated
June 29, 2012
Posted
September 17, 2012
Terminated
March 25, 2013
Address
2350 Qume Dr, San Jose, CA, 95131-1812

Description

BD FACS 7-color Setup beads BD Biosciences, San Jose, CA 95131. BD FACS 7 color Setup Beads are for in vitro diagnostic use on BD FACSCanto or BD FACSCanto II systems. The beads are run with BD FACSCanto clinical software. This system provides application specific setup, including automated compensation and cytometer quality control (QC) with minimal user intervention.

Reason

Incorrect version of the CD was included with BD FACS 7 Color Setup Beads.

Action

BD Biosciences sent a Important Product Recall Information letter beginning July 3, 2012, to all affected consignees via Fed Express. The letter identified the product the problem and the action needed to be taken by the customer. The customers were informed that the CD enclosed in their recent purchase of BD FACS 7-Color Setup Beads, lot number 33412, contain CD version 2.0 instead of the newer version 3.0. The letter goes on to explain the the potential risk and include a table of listed range values to determine if a potential false pass or false fail for the listed detectors has occurred. Customers are asked to check the CD with the kit to ensure that it is version 3.0 (the kit lot number ID is on the CD label). If it is not the correct version, discard version 2.0 and do not use it with this bead kit. Call BAD customers service for a replacement. If the 2.0 kit was already uses, letters request that clinicians inspect the FITC and APC-CY7 stains in the plots of the laboratory reports for the clinical applications in BD FACSCanto clinical software for compromised resolution between negative and dim populations. The verification form/effectiveness check should be completed and returned as requested. Questions or concerns may be directed to BD Customer Support at 1-855-236-2772.

Distribution

Worldwide Distribution -- USA (nationwide) including the countries of Australia, Belgium, Canada, Chile, Korea, Mexico, Singapore and Taiwan.

Quantity

96 kits