AD-TECH Spencer Probe Depth Electrode Product Usage: The Ad-Tech Depth Electrodes (Depth Electrodes, Foramen Ovale Depth Electrodes, Macro Micro Depth Electrodes, Spencer Probe Depth Electrodes, Wyler Sphenoidal Depth Electrodes) are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.
Recall
- Recall Number
- Z-2396-2019
- Event Number
- 82719
- Firm
- Ad-Tech Medical Instrument Corporation
- FEI Number
- 2183456
- Product Code
- GZL
- Status
- Terminated
- Root Cause
- Labeling Change Control
- Initiated
- April 16, 2019
- Terminated
- May 27, 2022
- Address
- 400 W Oakview Pkwy, Oak Creek, WI, 53154-7213
Description
AD-TECH Spencer Probe Depth Electrode Product Usage: The Ad-Tech Depth Electrodes (Depth Electrodes, Foramen Ovale Depth Electrodes, Macro Micro Depth Electrodes, Spencer Probe Depth Electrodes, Wyler Sphenoidal Depth Electrodes) are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.
An Ad-Tech Clinical Specialist, attended a case on April 15, 2019. During the case the surgeon encountered an issue after implanting four Ad-Tech depth electrodes. Upon initial device testing subsequent to implantation it was identified that two of the devices that were labeled as 8 contact depth electrodes (SD08R-SP05X-000) were in fact 6 contact depth electrodes (SD06R-SP05X-000) . There was no impact to the patient due to this incident.
Via a MEDICAL DEVICVE RECALL letter dated, April 17, 2019, Ad-Tech advised their consignees of the mix-up, requested that they check their stock, returned the MEDICAL DEVICE RECALL RETURN RESPONSE, and return the subject devices. Distributors were requested to conduct a sub-recall to their customers.
Worldwide - Nationwide distribution in the state of Illinois and countries of Russia and United Kingdom
18