FDA Recall Terminated

ES0102, AEM Disposable Curved Scissors, 3/4", 35cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical procedures

Recall: Z-2396-2011 · Initiated April 7, 2011

Recall

Recall Number
Z-2396-2011
Event Number
58518
Firm
Encision, Inc.
FEI Number
1000117361
Product Code
GEI
Status
Terminated
Root Cause
Package design/selection
Initiated
April 7, 2011
Posted
May 31, 2011
Terminated
August 9, 2011
Address
6797 Winchester Cir, Boulder, CO, 80301-3513

Description

ES0102, AEM Disposable Curved Scissors, 3/4", 35cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical procedures

Reason

Sterility may be compromised due to holes in packaging.

Action

Encision, Inc. sent an Important Safety Notice letter dated April 7, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to either carefully inspect any remaining pouches for readily visible holes in the Tyvek pouch, or return the affected lots to Encision for inspection. Customers were also asked to fill in and fax back the attached Confirmation to 303-444-2693.

Distribution

Worldwide Distribution - USA (nationwide) and the countries of Australia and New Zealand

Quantity

952 Ten-Count Boxes