FDA Recall Open, Classified

INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLE

Recall: Z-2392-2023 · Initiated June 22, 2023

Recall

Recall Number
Z-2392-2023
Event Number
92664
Firm
Canon Medical System, USA, INC.
FEI Number
2020563
Product Code
OWB
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
June 22, 2023
Posted
August 11, 2023
Address
2441 Michelle Dr, Tustin, CA, 92780-7047

Description

INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLE

Reason

The C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue using the operators console, the switches on the back of the column, switches on the side of the flat panel detector, and the emergency stop will continue to function normally.

Action

On 6/22/23 correction notices were emailed to customers who were advised of the following: If the table side control does not function then the procedure can continue using the operators console in the operator room, the switches on the back of the column and the switches on the side of the flat panel detector. The emergency stop should not be affected and will continue to function normally. Neither manual operation or auto positioning of the C-arm can be performed on the console during procedure, and the C-arm will not operate. A firm representative will contact you to schedule a time to replace the PWB on the console. Complete and return the customer reply form to [email protected] For questions, contact the InTouch team at [email protected] or your service representative at 800.521.1968

Distribution

US Nationwide Distribution: NY, CO, IL, CA, MD, MN, WA

Quantity

12