FDA Recall Terminated

Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087

Recall: Z-2392-2016 · Initiated June 30, 2015

Recall

Recall Number
Z-2392-2016
Event Number
74714
Firm
Hitachi Medical Systems America Inc
FEI Number
1528028
Product Code
JAK
Status
Terminated
Root Cause
Component design/selection
Initiated
June 30, 2015
Terminated
November 1, 2016
Address
1959 Summit Commerce Park, Twinsburg, OH, 44087-2371

Description

Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087

Reason

The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.

Action

Hitachi sent an Device Correction Notice dated April 1, 2015, to all affected customers. The letter identified the product the problem and the action to be taken by the customer. Customers were informed that there is a problem with the Hitachi Medical Systems America, Inc.CT system; the issue is with a power supply in the CT gantry sub-system. During fast rotation (i.e., 35 sec), the centrifugal force applied to the power supply may cause it to fail over time. For further questions, please call (330) 425-1313 ext. 3720.

Distribution

US Distribution to the states of : CA, FL, GA, IA, IL, IN, MD, NC, NY, OH, PA, SD, TN, TX & WY., and Internationally to Brazil and Mexico.

Quantity

33 units