Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087
Recall
- Recall Number
- Z-2392-2016
- Event Number
- 74714
- Firm
- Hitachi Medical Systems America Inc
- FEI Number
- 1528028
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- June 30, 2015
- Terminated
- November 1, 2016
- Address
- 1959 Summit Commerce Park, Twinsburg, OH, 44087-2371
Description
Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087
The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.
Hitachi sent an Device Correction Notice dated April 1, 2015, to all affected customers. The letter identified the product the problem and the action to be taken by the customer. Customers were informed that there is a problem with the Hitachi Medical Systems America, Inc.CT system; the issue is with a power supply in the CT gantry sub-system. During fast rotation (i.e., 35 sec), the centrifugal force applied to the power supply may cause it to fail over time. For further questions, please call (330) 425-1313 ext. 3720.
US Distribution to the states of : CA, FL, GA, IA, IL, IN, MD, NC, NY, OH, PA, SD, TN, TX & WY., and Internationally to Brazil and Mexico.
33 units