FDA Recall Open, Classified

Description/REF: RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-SPC; RA CATH KIT: 20 GA X 5 IN/ASK-04020-MW; ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM5; CATH SL 20 GA X 5 CM HYDROPHILIC/ASK-00520-JH; CATH S-L 20 GA X 5 CM HYDROPHILIC/ASK-00520-UHC; ARTERIAL LINE KIT: 20 GA x 5IN/ASK-00820-CL; RA CATH KIT: 20 GA X 8 CM X 12 CM/ASK-00820-WC; ARTERIAL CATHETERIZATION KIT/ASK-01218-HM; SAC KIT: 20 GA X 12CM/ASK-01220-HMH; ARTERIAL CATH KIT: 20GA X 12CM/ASK-01220-NWM; ARTERIAL CATH KIT: 20 GA X 12 CM/ASK-01220-UH; ARTERIAL CATH: 18 GA X 6 1/2IN/ASK-01618-AU; RA CATH KIT: 18 GA X 6 IN/ASK-01618-CL; ARTERIAL CATH KIT: 18 GA X 6 1/2 IN/ASK-01618-JH; ARTERIAL CATHETERIZATION KIT/ASK-01618-PH; ARTERIAL CATH KIT: 18 GA X 6 1/2IN/ASK-01618-RU; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018-HF; ARTERIAL CATH KIT: 18 GA X 6 IN (16 CM)/ASK-04018-NS; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-ST; RA CATH KIT: 20 GA X 1 1/2 IN (3.81 CM)/ASK-04120-BGH; ART/VESS CATH KIT:20GAX4.45CM/20GAX12CM/ASK-04500-SM; ARTERIAL LINE KIT/ASK-04500-UCH; ARTERIAL LINE KIT/ASK-04510-EUH2; ARTERIAL LINE KIT: 20GA X 5IN (12CM)/ASK-04510-MS4; ARTERIAL LINE KIT/ASK-04510-UPM; ARTERIAL CATH KIT: 20GA X 5CM/SAC-00520-1A; ARTERIAL CATH KIT: 20GA X 5CM/SAC-00520-X1A; ARTERIAL CATH KIT: 22GA X 5CM/SAC-00522-1A ARTERIAL CATH KIT: 22 GA X 5CM/SAC-00522-X1A; ARTERIAL CATH KIT: 20GA X 8CM/SAC-00820-1A; ARTERIAL CATH KIT: 20GA X 8CM/SAC-00820-X1A; SAC KIT: 18 GA X 12CM/SAC-01218-1A; SAC KIT: 18 GA X 12CM/SAC-01218-X1A; SAC KIT: 20 GA X 12CM/SAC-01220-1A; SAC KIT: 20 GA X 12CM/SAC-01220-X1A; SAC KIT: 18 GA X 16CM/SAC-01618-1A; SAC KIT: 18 GA X 16CM/SAC-01618-X1A; ACCESS TRAY/UM-04018; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-ABM; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-BGH; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-LOL; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-MHS; RA CATH KIT: 20 GA X 1 3/4IN (4.45 CM)/ASK-04020-MIB3; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-PMC1; RA CATH SET: 20 GA X 1 3/4 IN/ASK-04020-URMC; RA CATH KIT: 20 GA X 1 1/2 IN (3.81 CM)/ASK-04220-FMH1; RA CATH KIT: 20 GA X 4.45 CM/NA-04020-1A; RA CATH KIT: 20 GA X 4.45 CM W BIO/NA-04020-B; RA CATH KIT: 20 GA X 4.45 CM W CHG/NA-04020-C; VESSEL CATH KIT: 20 GA X 5IN (12 CM)/ASK-04550-NYP; CVC KIT: 16 GA X 8 IN (20 CM)/ASK-04300-PHP1; CVC KIT: 16 GA X 8 IN (20 CM)/ASK-04301-TJ; ARTERIAL ACCESS KIT/ASK-04001-VAMC1; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-WBH

Recall: Z-2391-2026 · Initiated April 21, 2026

Recall

Recall Number
Z-2391-2026
Event Number
98793
Firm
ARROW INTERNATIONAL, LLC
FEI Number
3013162291
Product Code
OFK
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
April 21, 2026
Posted
June 13, 2026
Address
3015 Carrington Mill Blvd, Morrisville, NC, 27560-5437

Description

Description/REF: RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-SPC; RA CATH KIT: 20 GA X 5 IN/ASK-04020-MW; ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM5; CATH SL 20 GA X 5 CM HYDROPHILIC/ASK-00520-JH; CATH S-L 20 GA X 5 CM HYDROPHILIC/ASK-00520-UHC; ARTERIAL LINE KIT: 20 GA x 5IN/ASK-00820-CL; RA CATH KIT: 20 GA X 8 CM X 12 CM/ASK-00820-WC; ARTERIAL CATHETERIZATION KIT/ASK-01218-HM; SAC KIT: 20 GA X 12CM/ASK-01220-HMH; ARTERIAL CATH KIT: 20GA X 12CM/ASK-01220-NWM; ARTERIAL CATH KIT: 20 GA X 12 CM/ASK-01220-UH; ARTERIAL CATH: 18 GA X 6 1/2IN/ASK-01618-AU; RA CATH KIT: 18 GA X 6 IN/ASK-01618-CL; ARTERIAL CATH KIT: 18 GA X 6 1/2 IN/ASK-01618-JH; ARTERIAL CATHETERIZATION KIT/ASK-01618-PH; ARTERIAL CATH KIT: 18 GA X 6 1/2IN/ASK-01618-RU; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018-HF; ARTERIAL CATH KIT: 18 GA X 6 IN (16 CM)/ASK-04018-NS; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-ST; RA CATH KIT: 20 GA X 1 1/2 IN (3.81 CM)/ASK-04120-BGH; ART/VESS CATH KIT:20GAX4.45CM/20GAX12CM/ASK-04500-SM; ARTERIAL LINE KIT/ASK-04500-UCH; ARTERIAL LINE KIT/ASK-04510-EUH2; ARTERIAL LINE KIT: 20GA X 5IN (12CM)/ASK-04510-MS4; ARTERIAL LINE KIT/ASK-04510-UPM; ARTERIAL CATH KIT: 20GA X 5CM/SAC-00520-1A; ARTERIAL CATH KIT: 20GA X 5CM/SAC-00520-X1A; ARTERIAL CATH KIT: 22GA X 5CM/SAC-00522-1A ARTERIAL CATH KIT: 22 GA X 5CM/SAC-00522-X1A; ARTERIAL CATH KIT: 20GA X 8CM/SAC-00820-1A; ARTERIAL CATH KIT: 20GA X 8CM/SAC-00820-X1A; SAC KIT: 18 GA X 12CM/SAC-01218-1A; SAC KIT: 18 GA X 12CM/SAC-01218-X1A; SAC KIT: 20 GA X 12CM/SAC-01220-1A; SAC KIT: 20 GA X 12CM/SAC-01220-X1A; SAC KIT: 18 GA X 16CM/SAC-01618-1A; SAC KIT: 18 GA X 16CM/SAC-01618-X1A; ACCESS TRAY/UM-04018; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-ABM; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-BGH; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-LOL; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-MHS; RA CATH KIT: 20 GA X 1 3/4IN (4.45 CM)/ASK-04020-MIB3; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-PMC1; RA CATH SET: 20 GA X 1 3/4 IN/ASK-04020-URMC; RA CATH KIT: 20 GA X 1 1/2 IN (3.81 CM)/ASK-04220-FMH1; RA CATH KIT: 20 GA X 4.45 CM/NA-04020-1A; RA CATH KIT: 20 GA X 4.45 CM W BIO/NA-04020-B; RA CATH KIT: 20 GA X 4.45 CM W CHG/NA-04020-C; VESSEL CATH KIT: 20 GA X 5IN (12 CM)/ASK-04550-NYP; CVC KIT: 16 GA X 8 IN (20 CM)/ASK-04300-PHP1; CVC KIT: 16 GA X 8 IN (20 CM)/ASK-04301-TJ; ARTERIAL ACCESS KIT/ASK-04001-VAMC1; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-WBH

Reason

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

Action

On 4/21/2026, correction notices (Firm Event Number: EIF-000608, lidocaine and bupivacaine) were mailed to customers who were asked to do the following: Arrow International LLC is voluntarily recalling product due to the Huons Co. Ltd. recall of Lidocaine and Bupivacaine, which is included in impacted Arrow Kits & Sets. Medical facilities 1. Provide this correction notice to all persons who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2. Check your inventory for product within the scope of this Medical Device Notification and place a copy of this notice with all affected product. 3. Complete and return the response form virtually via: https://fca.myteleflex.com/en/recall/000608 Distributors: 1. Provide Medical Device Notification to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. 2. Check your inventory for affected product. Place a copy of this notification with all affected product and ensure that the notification is included with future distributions of affected product. 3. Complete and return the response form via link: https://fca.myteleflex.com/en/recall/000608 4. If you have further distributed product outside of your country, please notify Teleflex Customer Service by return email. If you have questions contact firm's Customer Service: Link: https://fca.myteleflex.com/en/recall/000608, Customer Service: Telephone: 1-866-396-2111, FAX: 1-855-419-8507 Email: [email protected] On 5/29/2026, additional amended notices (Firm Event Number: EIF-000608-01) were mailed to include additional affected devices. On 5/1/2026, additional notices (Firm Event Number: EIF-000609, sodium chloride) were mailed and emailed to customers who have 0.9% sodium chloride (10mL saline solution ampule) in kits and sets.

Distribution

US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.

Quantity

491,489