FDA Recall Terminated

Marco Shoulder Stabilization Kit, Product Code 711100, (previous product code 100601), Sample Kit S711100

Recall: Z-2389-2021 · Initiated July 23, 2021

Recall

Recall Number
Z-2389-2021
Event Number
88476
Firm
Aspen Surgical Products, Inc.
FEI Number
3002750084
Product Code
CCX
Status
Terminated
Root Cause
Process control
Initiated
July 23, 2021
Terminated
July 26, 2022
Address
6945 Southbelt Dr SE, Caledonia, MI, 49316-7664

Description

Marco Shoulder Stabilization Kit, Product Code 711100, (previous product code 100601), Sample Kit S711100

Reason

Affected products were manufactured with materials found to have a bioburden level exceeding the limit at which the product's sterilization process was validated. As a result, the firm cannot provide full assurance that the affected products were effectively sterilized, and the product may compromise a sterile field and increase the risk of a surgical site infection.

Action

Required Actions: 1. Check all storage and usage locations to confirm if you have any units of the affected item codes in your possession. Please note that these products bear an expiration date; any expired product should be discarded regardless of whether they are affected by this corrective action. 2. Distributors: Please check your inventory for affected products as indicated above. For products that you have distributed, please forward this notification to your customers who have received affected product. Please also provide to Aspen a list of customers who have received these products, including the number of each product and contact information for each customer, to [email protected]. 3. Review, complete, sign and return the Acknowledgement Form which accompanies the notice.. 4. Destroy and discard any affected product found in your inventory. Your sales representative can assist with disposal of affected product if needed. 5. Contact Aspen Customer Service, at 888-364-7004 or customerservice@aspensurgical for information on credit options. At this time, we expect that replacement product will be available by October, 2021; however, we hope to expedite production and will contact you with updates on supply availability. In the interim, until Aspen suggests considering the following alternatives: o o Marco Shoulder Stabilization Kit: Smith & Nephew SPIDER Shoulder Stabilization Kit (7210573) o o Jessie Arm Suspension Kit: McConnell Arm Support (12-401) 6. Share this notice with others in your facility who need to be aware of this removal. Contact other facilities associated with your organization that may have received units of affected lots. 7. Maintain awareness of this notice until all affected product has been returned to Aspen Surgical or destroyed. 8. Keep a copy of this notice with affected product until returned to Aspen.

Distribution

US nationwide distribution. No foreign distribution.

Quantity

4270