NicoletOne photic adapter cable 085-463700
Recall
- Recall Number
- Z-2389-2008
- Event Number
- 48758
- Firm
- Cardinal Health NeuroCare Division
- FEI Number
- 3010611950
- Product Code
- GWQ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- June 5, 2008
- Posted
- September 21, 2008
- Terminated
- October 13, 2010
- Address
- 5225 Verona Rd, Madison, WI, 53711-4497
Description
NicoletOne photic adapter cable 085-463700
Cardinal Health NeuroCare is implementing a voluntary field correction to replace a quantity of photic adapter cables, part number 085-463700, used to connect the NicLED Photic Strobe option, part number 698-625500, with NicoletOne EEG system. The photic adapter cable has a wiring error which delivers a reduced voltage and results in a decreased intensity of the NicLED photic. This reduced intensi
A "Cardinal Health Urgent: Medical Device Correction" letter was sent to consignees. The letter described the product and the problem. The consignees were asked to confirm completion of the cable exchange by returning an information card. If you have questions, contact Cardinal Health (formerly VIASYS) NeuroCare Technical Support Group at 1-800-356-0007 (US), 608-273-5000 (OUS) or email [email protected].