FDA Recall Terminated

NicoletOne photic adapter cable 085-463700

Recall: Z-2389-2008 · Initiated June 5, 2008

Recall

Recall Number
Z-2389-2008
Event Number
48758
Firm
Cardinal Health NeuroCare Division
FEI Number
3010611950
Product Code
GWQ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
June 5, 2008
Posted
September 21, 2008
Terminated
October 13, 2010
Address
5225 Verona Rd, Madison, WI, 53711-4497

Description

NicoletOne photic adapter cable 085-463700

Reason

Cardinal Health NeuroCare is implementing a voluntary field correction to replace a quantity of photic adapter cables, part number 085-463700, used to connect the NicLED Photic Strobe option, part number 698-625500, with NicoletOne EEG system. The photic adapter cable has a wiring error which delivers a reduced voltage and results in a decreased intensity of the NicLED photic. This reduced intensi

Action

A "Cardinal Health Urgent: Medical Device Correction" letter was sent to consignees. The letter described the product and the problem. The consignees were asked to confirm completion of the cable exchange by returning an information card. If you have questions, contact Cardinal Health (formerly VIASYS) NeuroCare Technical Support Group at 1-800-356-0007 (US), 608-273-5000 (OUS) or email [email protected].

Distribution

CA FL GA ID IN LA MD ME MI MO NC NY PA TN TX WA WI China Denmark Indonesia Japan Korea Mexico Netherlands Russia Saudi Arabia Slovenia

Quantity

92