FDA Recall Terminated

Instrumentation Laboratory ACL TOP 700 PN 0000280010 ACL TOP is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters

Recall: Z-2388-2012 · Initiated August 3, 2012

Recall

Recall Number
Z-2388-2012
Event Number
62894
Firm
Instrumentation Laboratory Co.
FEI Number
1217183
Product Code
GKP
Status
Terminated
Root Cause
Software design
Initiated
August 3, 2012
Posted
September 14, 2012
Terminated
October 2, 2015
Address
180 Hartwell Road, Bedford, MA, 01730-2443

Description

Instrumentation Laboratory ACL TOP 700 PN 0000280010 ACL TOP is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters

Reason

Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low

Action

The firm, Instrumentation Laboratory, sent a "Mandatory Test Parameter Upgrade" letter dated June 7, 2012 with upgrade kits P-14.8 via USPS on August 3 thru August 10 to US and Canadian customers and notified International Affiliates via e-mail on August 2, 2012. The customers were ask to complete and return the Mandatory Test Parameter Upgrade Tracking Form via fax: 781-861-4207 or email: [email protected]. If you have any questions, call 781-861-4204.

Distribution

Worldwide distribution: USA (nationwide) and countries of: ARGENTINA, AUSTRALIA, AUSTRIA, BELARUS, BELGIUM, BRAZIL, BRUNEI DARUSSALAM, BULGARIA, CHILE, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF LUXEMBOURG, MACAU, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PANAMA, PERU, POLAND, PORTUGAL, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SLOVAKIA (Slovak Republic), SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, PROVINCE OF CHINA, THAILAND, TURKEY, UNITED KINGDOM, URUGUAY and VIETNAM.

Quantity

304 units