Infusomat SPACE PUMP IV SET, 2 CARESITES, ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 363420
Recall
- Recall Number
- Z-2387-2024
- Event Number
- 94823
- Firm
- B. Braun Medical, Inc.
- FEI Number
- 2523676
- Product Code
- FPA
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- June 17, 2024
- Posted
- July 17, 2024
- Address
- 901 Marcon Blvd, Allentown, PA, 18109-9512
Description
Infusomat SPACE PUMP IV SET, 2 CARESITES, ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 363420
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
B. Braun Medical Inc. (BBMI) is issuing a voluntary Urgent Medical Device Correction letter via US Postal Service Certified Mail with registered return receipt on 6/17/24. Letter states reason for recall, health risk and action to take: Corrective actions are being implemented to prevent these concerns in future manufacturing lots through optimization of the disks that provide the sealing function within these valves. Until such time as lots with these corrective actions are available, users are asked to follow the "Interim Measures for Users." A follow-up communication will be provided to users when lots with these corrective actions are in place and the interim measures can be stopped. At this time, we anticipate that corrected sets may become available 04 2024, but more definitive information will be provided in our follow-up notice. Interim Measures for Users: 1. Until further notice, when administering secondary medications via piggyback please clamp the primary line above the upper Y site using the available slide clamp on the pump administration set. This will require unclamping of the primary line after delivery of the secondary medication. The instructions for this differ depending on the type of pump and software as outlined below : a. Attachment 1 provides workflow recommendations for lnfusomat Space Pumps with "U" software versions. b. Attachment 2 provides workflow recommendations for lnfusomat Space Pumps with "G" software versions and "L" (Canada only) software versions. c. Attachment 3 provides workflow recommendations for Outlook ES Pumps. Note: This does not change workflows for rates of infusion greater than 300 ml/hr for lnfusomat Pump Administration Sets and 125 ml/hr with Outlook Pump Administration sets where clamping of the primary infusion line is always required per the product IFU to prevent concurrent flow. 2. Inform all users and post the applicable attachment(s) in care areas where secondary infusions may be administered and where pu
Worldwide distribution - US Nationwide and the country of Canada.
68205 units