FDA Recall Terminated

Access and Access 2 Immunoassay System

Recall: Z-2384-2008 · Initiated April 3, 2008

Recall

Recall Number
Z-2384-2008
Event Number
48186
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
CGN
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 3, 2008
Posted
September 21, 2008
Terminated
November 29, 2011
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

Access and Access 2 Immunoassay System

Reason

The field action was initiated after the firm confirmed reports of premature failure of the waste pump tubing in the Access and Access 2 Immunoassay Systems. Premature failure of the waste pump tubing may negatively affect precision.

Action

A Product Corrective Action (PCA) letter with attached Fax-Back Customer Response form was mailed on April 16, 2008 via US Postal mail to all affected customers with Access and Access 2 Immunoassay Systems, Synchron LXi 725 Systems and UniCel DxC 600i Systems. Customers were informed about the issue and asked to inspect for leaking or damaged tubing on all peristatic pump Mfg. No. 01-00772205 with a SN higher than 2234980. A toll free Customer Technical Support telephone was included in the letter. After further investigation by the firm, an Updated PCA letter with attached Fax-Back Customer Response form was mailed on June 30, 2008 via US Postal mail to include all customers with the following products: UniCel Dxl 800 and DxC 880i Systems with pump manufacturing number (Mfg.) of 01-010341-1 and serial number (SN) 22454201 and above, and UniCel Dxl 600 Systems with a pump Mfg. No. A24468 and SN of 22965501. Customers were informed about the issue and asked to inspect for leaking or damaged tubing on all peristatic pump SN referenced. If you have questions, contact Customer Technical Support at 1-800-854-3633.

Distribution

Nationwide, and in Canada

Quantity

694 Worldwide; 656 in US & 38 in Canada total for all products