VERTIER Surgical Table, Catalog #0788-100-001S (SIDNE-enabled standard table with 5th wheel); distributed by Stryker Communications, Flower Mound, TX. Is an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system.
Recall
- Recall Number
- Z-2379-2008
- Event Number
- 49021
- Firm
- Stryker Communications Corp
- FEI Number
- 3002975342
- Product Code
- FQO
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- July 2, 2008
- Posted
- September 18, 2008
- Terminated
- January 7, 2010
- Address
- 1410 Lakeside Pkwy, Ste 100, Flower Mound, TX, 75028-4026
Description
VERTIER Surgical Table, Catalog #0788-100-001S (SIDNE-enabled standard table with 5th wheel); distributed by Stryker Communications, Flower Mound, TX. Is an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system.
Hydraulic lines responsible for tilting the surgical table have the potential to be severed when articulating table to its lowest position, possibly resulting in the unexpected and rapid movement of the table.
Urgent: Device Recall letters were sent to consignees on 07/02/08 via FedEx with return receipt. Consignees were asked to discontinue use of the tables until a Stryker representative has come to the facility and updated the table. The firm has identified and validated replacement parts that prevent this occurrence.
Nationwide Distribution --- including states of AR, CA, FL, GA, ID, IL, IN, KS, KY, MN, MO, ND, NJ, NY, OH, OK, OR, SD, TX, UT, VA and Washington DC.
4 units.