FDA Recall Terminated

VERTIER Surgical Table, Catalog #0788-100-001 (Standard table with 5th wheel); distributed by Stryker Communications, Flower Mound, TX. Is an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system.

Recall: Z-2377-2008 · Initiated July 2, 2008

Recall

Recall Number
Z-2377-2008
Event Number
49021
Firm
Stryker Communications Corp
FEI Number
3002975342
Product Code
FQO
Status
Terminated
Root Cause
Device Design
Initiated
July 2, 2008
Posted
September 18, 2008
Terminated
January 7, 2010
Address
1410 Lakeside Pkwy, Ste 100, Flower Mound, TX, 75028-4026

Description

VERTIER Surgical Table, Catalog #0788-100-001 (Standard table with 5th wheel); distributed by Stryker Communications, Flower Mound, TX. Is an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system.

Reason

Hydraulic lines responsible for tilting the surgical table have the potential to be severed when articulating table to its lowest position, possibly resulting in the unexpected and rapid movement of the table.

Action

Urgent: Device Recall letters were sent to consignees on 07/02/08 via FedEx with return receipt. Consignees were asked to discontinue use of the tables until a Stryker representative has come to the facility and updated the table. The firm has identified and validated replacement parts that prevent this occurrence.

Distribution

Nationwide Distribution --- including states of AR, CA, FL, GA, ID, IL, IN, KS, KY, MN, MO, ND, NJ, NY, OH, OK, OR, SD, TX, UT, VA and Washington DC.

Quantity

16 units.