FDA Recall Terminated

CARTO 3 V1.05(FG-5400-00). The CARTO 3 V1.05 System is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps and cardiac chamber geometry maps. The acquired patient signals, including body surface ECG and intracardiac electrograms may also be displayed in real time on the System's display screen. The CARTO 3 V1.05 System is also intended to support EP procedures, maintaining CARTO XP System capabilities, in the presence of a high metallic environment and provide a data communication channel to the Stereotaxis Niobe Catheter Guidance System as well as in a regular EP lab. The CARTO 3 V1.05 System includes the CartoMerge capability to import, register and merge CT or MRI structural images with CARTO Maps physiological information and real time catheter navigation, the Fast Anatomical Mapping (FAM) functionality, enabling creation of cardiac maps using'catheters with magnetic proximal and distal location sensors and the CartoSound image integration functionality, that enables integration of intracardiac echo (ICE) visualization to provide 3D combined maps. In addition to the use of specialized navigation catheters, the system is also intended for use with non-navigational electrophysiology catheters.

Recall: Z-2369-2012 · Initiated November 17, 2009

Recall

Recall Number
Z-2369-2012
Event Number
54659
Firm
Biosense Webster, Inc.
FEI Number
3003113629
Product Code
DQK
Status
Terminated
Root Cause
Labeling mix-ups
Initiated
November 17, 2009
Posted
September 12, 2012
Terminated
September 12, 2012
Address
15715 Arrow Hwy, Irwindale, CA, 91706-2006

Description

CARTO 3 V1.05(FG-5400-00). The CARTO 3 V1.05 System is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps and cardiac chamber geometry maps. The acquired patient signals, including body surface ECG and intracardiac electrograms may also be displayed in real time on the System's display screen. The CARTO 3 V1.05 System is also intended to support EP procedures, maintaining CARTO XP System capabilities, in the presence of a high metallic environment and provide a data communication channel to the Stereotaxis Niobe Catheter Guidance System as well as in a regular EP lab. The CARTO 3 V1.05 System includes the CartoMerge capability to import, register and merge CT or MRI structural images with CARTO Maps physiological information and real time catheter navigation, the Fast Anatomical Mapping (FAM) functionality, enabling creation of cardiac maps using'catheters with magnetic proximal and distal location sensors and the CartoSound image integration functionality, that enables integration of intracardiac echo (ICE) visualization to provide 3D combined maps. In addition to the use of specialized navigation catheters, the system is also intended for use with non-navigational electrophysiology catheters.

Reason

The recall was initiated because Biosense Webster has become aware of information regarding a potential hazard that may occur during the operation of a Bloom EP Stimulator device in conjunction with RF generators and other electrophysiology (EP) equipment, including the CARTO 3 Electropysiology Mapping System.

Action

Biosense Webster sent an " URGENT FIELD SAFETY NOTICE" dated November 20, 2009 to all affected customers who purchased the Carto 3 Electrophysiology Mapping System. The letter informed the customers of the problem identified and the actions to be taken. Customers were instructed to continue use of the CARTO 3 System and that it was safe and no further action is needed. Customers with questions about the notice were instructed to contact Biosense Webster representatives. On January 12, 2010 another letter was sent from Biosense to all their customers recommending to avoid any connection of an EP stimulator's pacing pins to the direct ports on the CARTO 3 System when applying RF energy, unless the set up is indicated by the stimulator's instructions for use. The letter also informed the customer that the user manual for the CARTO 3 System has been formally revised to include a warning. Customers were instructed to complete an acknowledgement form to confirm receipt of the revised user manual.

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Denmark, France, Germany, Israel, Italy, Netherlands, and United Kingdom.

Quantity

71 units total (30 units in US)