FDA Recall Terminated

TransWarmer Warming Infant Transport Mattress, Model no. 20421 Product Usage: The TransWarmer Warming Infant Transport Mattress is gel-filled, disposable mattress that provides heat for up to two hours when cold stress is a concern. The patented WarmGel heats up to 104 (when starting from 75 ) in 60 seconds or less. The product is sold in boxes of 6 Infant Transport Mattresses per box. The intended use of the product is to provide warmth during infant transport within the hospital or between hospitals.

Recall: Z-2364-2019 · Initiated July 8, 2019

Recall

Recall Number
Z-2364-2019
Event Number
83110
Firm
CooperSurgical, Inc.
FEI Number
1216677
Product Code
IMD
Status
Terminated
Root Cause
Error in labeling
Initiated
July 8, 2019
Terminated
May 28, 2020
Address
95 Corporate Dr, Trumbull, CT, 06611-1350

Description

TransWarmer Warming Infant Transport Mattress, Model no. 20421 Product Usage: The TransWarmer Warming Infant Transport Mattress is gel-filled, disposable mattress that provides heat for up to two hours when cold stress is a concern. The patented WarmGel heats up to 104 (when starting from 75 ) in 60 seconds or less. The product is sold in boxes of 6 Infant Transport Mattresses per box. The intended use of the product is to provide warmth during infant transport within the hospital or between hospitals.

Reason

The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is prohibited and could lead to serious health consequences, such as skin burns. some products do not contain the updated IFU.

Action

On July 8, 2019, the firm began notifying customers of the recall via Urgent Medical Device Correction letters. The letters informed customers that CooperSurgical has updated the IFU to clarify that use of the Transwarmer Mattress with other heat producing devices, such as an incubator, is prohibited. Such use could lead to serious health consequences, such as skin burns. Customers were provided with an updated version of the IFU. Customers were asked to complete and return the acknowledgement and receipt form to the firm. If you have any questions about this recall, please contact CooperSurgical at 203-601-5200 ext. 3300.

Distribution

Worldwide Distribution - US Nationwide in all 50 states. International distribution to Australia, Brazil, Canada, Ecuador, Germany, Gibraltar, India, Ireland, Kuwait, Macedonia, New Zealand, Oman, Poland, Puerto Rico, Qatar, Romania, Saudi Arabia, South Korea, Switzerland, United Arab Emirates, & United Kingdom.

Quantity

49813