FDA Recall Terminated

ReUnion TSA Shoulder System, Standard SR Humeral Head Trial, 48X18, Catalog number 5901-S-4818. Orthopedic manual surgical instrument. Howmedica Osteonics Corp. A Subsidiary of Stryker Corp. Stryker France Non-sterile Made in Cork Ireland

Recall: Z-2363-2015 · Initiated May 12, 2015

Recall

Recall Number
Z-2363-2015
Event Number
71445
Firm
Stryker Howmedica Osteonics Corp.
FEI Number
3003070421
Product Code
KWS
Status
Terminated
Root Cause
Process control
Initiated
May 12, 2015
Posted
August 10, 2015
Terminated
October 5, 2016
Address
325 Corporate Dr, Mahwah, NJ, 07430-2006

Description

ReUnion TSA Shoulder System, Standard SR Humeral Head Trial, 48X18, Catalog number 5901-S-4818. Orthopedic manual surgical instrument. Howmedica Osteonics Corp. A Subsidiary of Stryker Corp. Stryker France Non-sterile Made in Cork Ireland

Reason

A package labeled as Part #5901-S-4818, Lot MAC7C14, standard humeral head trial, contained part # 5901-E-4818 Lot MAC7C14, offset humeral head trial.

Action

Stryker Orthopaedics notified their branches/agencies via email on 5/12/2015 and sent notification letters and product accountability forms to hospital risk management/branches on 7/15/2015 and 7/16/2015

Distribution

US Nationwide

Quantity

21 units