FDA Recall Open, Classified

Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.

Recall: Z-2358-2026 · Initiated May 8, 2026

Recall

Recall Number
Z-2358-2026
Event Number
99036
Firm
Datex-Ohmeda Inc
FEI Number
3036470679
Product Code
FMT
Status
Open, Classified
Root Cause
Process control
Initiated
May 8, 2026
Posted
June 18, 2026
Address
3114 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.

Reason

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

Action

Firm began notifying customers on 5/8/2026 via letters titled "URGENT MEDICAL DEVICE CORRECTION." Customers were instructed to identify affected devices by following instructions provided in the letter. Devices with affected blenders should be removed from patient use, and the customer should contact a GE HealthCare Service Representative.

Distribution

Pending

Quantity

550 (367 US; 183 OUS)