FDA Recall Open, Classified

Microdot Xtra blood glucose test strips

Recall: Z-2357-2019 · Initiated July 26, 2018

Recall

Recall Number
Z-2357-2019
Event Number
82867
FEI Number
3004176168
Product Code
NBW
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
July 26, 2018
Address
Cambridge Sensors Limited Units 9 & 10 Cardinal Park Huntingdon United Kingdom

Description

Microdot Xtra blood glucose test strips

Reason

The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.

Action

On July 26, 2018, the firm sent an email to it's US distributor that requested a sub-recall.

Distribution

Illinois

Quantity

17,784 (USA)