FDA Recall Terminated

Alere INRatio Replacement Monitor Model Number: 0100137 Usage: The INRatio 2 PT Monitoring system is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The INRatio 2 PT Monitoring system is intended for use outside the body (in vitro diagnostic use) by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The INRatio 2 PT Monitoring System is not intended to be used for screening purpose.

Recall: Z-2356-2016 · Initiated July 11, 2016

Recall

Recall Number
Z-2356-2016
Event Number
74665
Firm
Alere San Diego, Inc.
FEI Number
1000125596
Product Code
GJS
Status
Terminated
Root Cause
Other
Initiated
July 11, 2016
Posted
August 19, 2016
Terminated
July 20, 2017
Address
9975 Summers Ridge Rd, San Diego, CA, 92121-2997

Description

Alere INRatio Replacement Monitor Model Number: 0100137 Usage: The INRatio 2 PT Monitoring system is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The INRatio 2 PT Monitoring system is intended for use outside the body (in vitro diagnostic use) by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The INRatio 2 PT Monitoring System is not intended to be used for screening purpose.

Reason

Alere San Diego is initiating a voluntary recall for the Alere INRatio/INRatio2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio/INRatio2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Action

Alere issued a press release on July 11, 2016. An Urgent Medical Device Letter dated July 26, 2016 was sent to affected customers on July 29, 2016. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to consult with their healthcare provider as soon as possible to transition to an alternate method of PT/INR testing and to contact the firm regarding return or disposal of unused product and obtaining device replacement from a different manufacturer. For questions contact the Alere INRatid1 Recall Hotline at 1-866-723-2535.

Distribution

Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam.