A LINK Z INTERVERTEBRAL BODY FUSION DEVICE Product Description: A Link Z Ti Unitary Plate-Cage 32x21x15 7 A Link Z Ti Unitary Plate-Cage 32x21x15 12 A Link Z Ti Unitary Plate-Cage 32x21x15 16 A Link Z Ti Unitary Plate-Cage 32x21x15 16 A Link Z Ti Unitary Plate-Cage 32x21x15 16 A Link Z Ti Unitary Plate-Cage 32x21x15 20 A Link Z Ti Unitary Plate-Cage 32x21x15 20 A Link Z Ti Unitary Plate-Cage 32x21x15 20 A Link Z Ti Unitary Plate-Cage 32x21x17 12 A Link Z Ti Unitary Plate-Cage 32x21x17 16 A Link Z Ti Unitary Plate-Cage 32x21x17 20 A Link Z Ti Unitary Plate-Cage 32x21x19 20 A Link Z Ti Unitary Plate-Cage 32x24x15 7 A Link Z Ti Unitary Plate-Cage 32x24x15 16 A Link Z Ti Unitary Plate-Cage 36x26x15 12 A Link Z Ti Unitary Plate-Cage 36x26x15 12 A Link Z Ti Unitary Plate-Cage 36x26x15 16 A Link Z Ti Unitary Plate-Cage 36x26x15 16 A Link Z Ti Unitary Plate-Cage 36x26x15 16 A Link Z Ti Unitary Plate-Cage 36x26x15 16 A Link Z Ti Unitary Plate-Cage 36x26x15 20 A Link Z Ti Unitary Plate-Cage 36x26x15 20 A Link Z Ti Unitary Plate-Cage 36x26x15 20 A Link Z Ti Unitary Open Interbody 32x24x12 7 A Link Z Ti Unitary Open Interbody 32x24x12 12 A Link Z Ti Unitary Open Interbody 32x24x16 7 A Link Z Ti Unitary Open Interbody 32x24x16 12 Product Usage: is a spinal device that is implanted in the intervertebral body space via an anterior approach to improve stability of the spine while supporting fusion.
Recall
- Recall Number
- Z-2354-2020
- Event Number
- 85642
- Firm
- Acuity Surgical Devices, LLC
- FEI Number
- 3010866843
- Product Code
- OVD
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- January 7, 2019
- Address
- 8710 N Royal Ln, Irving, TX, 75063-2539
Description
A LINK Z INTERVERTEBRAL BODY FUSION DEVICE Product Description: A Link Z Ti Unitary Plate-Cage 32x21x15 7 A Link Z Ti Unitary Plate-Cage 32x21x15 12 A Link Z Ti Unitary Plate-Cage 32x21x15 16 A Link Z Ti Unitary Plate-Cage 32x21x15 16 A Link Z Ti Unitary Plate-Cage 32x21x15 16 A Link Z Ti Unitary Plate-Cage 32x21x15 20 A Link Z Ti Unitary Plate-Cage 32x21x15 20 A Link Z Ti Unitary Plate-Cage 32x21x15 20 A Link Z Ti Unitary Plate-Cage 32x21x17 12 A Link Z Ti Unitary Plate-Cage 32x21x17 16 A Link Z Ti Unitary Plate-Cage 32x21x17 20 A Link Z Ti Unitary Plate-Cage 32x21x19 20 A Link Z Ti Unitary Plate-Cage 32x24x15 7 A Link Z Ti Unitary Plate-Cage 32x24x15 16 A Link Z Ti Unitary Plate-Cage 36x26x15 12 A Link Z Ti Unitary Plate-Cage 36x26x15 12 A Link Z Ti Unitary Plate-Cage 36x26x15 16 A Link Z Ti Unitary Plate-Cage 36x26x15 16 A Link Z Ti Unitary Plate-Cage 36x26x15 16 A Link Z Ti Unitary Plate-Cage 36x26x15 16 A Link Z Ti Unitary Plate-Cage 36x26x15 20 A Link Z Ti Unitary Plate-Cage 36x26x15 20 A Link Z Ti Unitary Plate-Cage 36x26x15 20 A Link Z Ti Unitary Open Interbody 32x24x12 7 A Link Z Ti Unitary Open Interbody 32x24x12 12 A Link Z Ti Unitary Open Interbody 32x24x16 7 A Link Z Ti Unitary Open Interbody 32x24x16 12 Product Usage: is a spinal device that is implanted in the intervertebral body space via an anterior approach to improve stability of the spine while supporting fusion.
Due to a manufacturing process issue, it allowed vertebral body bone screws to not seat properly within the spinal device/cage.
On 12/14/2018, the firm notified its supplier via email requesting to visually review its inventory which determined that some screw holes were oval shaped. On 7/23/2019, all inventory was quarantined as part of the removal and was scrapped from inventory. Due to the low risk to patients and low probability, the firm ownly initially begin removal of product from internal locations (main warehouse) for the purpose of inspection and if necessary, repair or destruction. Removal Summary No additional customer or distributor communication took place during the removal.
US Nationwide distribution including in the states of TX, FL, NC, LA, KY, WI, SD, IL, CA, and KS. O.U.S.: None
A total of 27 lots producing 178 units