FDA Recall Open, Classified

Medtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pump revision kit)

Recall: Z-2352-2024 · Initiated May 10, 2024

Recall

Recall Number
Z-2352-2024
Event Number
94779
Firm
Medtronic Neuromodulation
FEI Number
2182207
Product Code
LKK
Status
Open, Classified
Root Cause
Device Design
Initiated
May 10, 2024
Posted
July 12, 2024
Address
7000 Central Ave Ne, Minneapolis, MN, 55432-3568

Description

Medtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pump revision kit)

Reason

Design update to the Ascenda Intrathecal Catheter Models 8780, 8781, and 8784. The intent of the design update is to reduce the potential for tissue growth into the Ascenda catheter connector which may potentially lead to catheter occlusion.

Action

Medtronic issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 05/10/2024 electronically (DocuSign) to each listed physician. For Physicians that do not have an email address or a DocuSign account, the Consignee Notification and Customer Confirmation Form were sent via mail UPS/USPS. The notice explained the issue, provided clinician and patient recommendations, and requested that the notice be shared with all those to whom the devices was transferred. If you have any questions regarding this communication, please contact Medtronic Technical Services at 1-800-707-0933 or your local Medtronic representative. On 03/12/2025, Medtronic issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees via mail. The Consignee Notifications informs the customer of availability of Ascenda catheters with the new design and requests the return of Ascenda catheters that were manufactured prior to the design update of 05/10/2024, and requested that the notice be shared with all those to whom the devices was transferred. For questions, please contact Medtronic Technical Services at 1-800-707-0933 or your local Medtronic representative.

Distribution

Worldwide Distribution: US (nationwide) and OUS (foreign) countries of: Argentina, Aruba, Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, Colombia, Cyprus, Czech Rep, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Guadeloupe, Germany, Greece, Hong Kong, Hungary, Indonesia, Italy, Iran, Islamic Republic, Iraq, Ireland, Jordan, Kazakhstan, Korean Republic of, Kuwait, Latvia, Lebanon, Lithuania, Libya, Luxembourg, Malaysia, Malta, Mexico, Morocco, Netherland, New Zealand, Norway, Oman, Panama, Pakistan, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Switzerland, Sweden, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirate, and United Kingdom.

Quantity

111762 units