Cascade IOMAX Cortical Module, REF: 190296-200
Recall
- Recall Number
- Z-2349-2024
- Event Number
- 94764
- Firm
- Cadwell Industries Inc
- FEI Number
- 3020018
- Product Code
- GWF
- Status
- Open, Classified
- Root Cause
- Process design
- Initiated
- May 17, 2024
- Posted
- July 12, 2024
- Address
- 909 N Kellogg St, Kennewick, WA, 99336-7669
Description
Cascade IOMAX Cortical Module, REF: 190296-200
Electroneurodiagnostic medical device may contain a defective printed circuit board which may cause damage to the other internal components resulting in unintended electrical stimulation.
On 5/17/2024, Cadwell Industries, Inc. issued "Urgent: Medical Device Recall" notification to affected consignees via E-Mail. Cadwell ask consignees to take the following actions: 1. Cease use of the device immediately and return all identified devices to Cadwell Industries, Inc. 2. Please complete the attached response form and return via email to [email protected]. 3. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred the IOMAX Cortical Module to. 4. Please complete acknowledgement and product replacement form.
US: IL CA OUS: Canada, Germany, India, Italy, Mexico, Poland, South Africa, Spain, Taiwan, United Kingdom
26 units