FDA Recall Terminated

Picasso/Picasso Lite lasers Intended use: Dentistry.

Recall: Z-2349-2010 · Initiated April 15, 2010

Recall

Recall Number
Z-2349-2010
Event Number
56330
Firm
AMD Lasers, LLC
FEI Number
3003258642
Product Code
GEX
Status
Terminated
Root Cause
Radiation Control for Health and Safety Act
Initiated
April 15, 2010
Posted
September 10, 2010
Terminated
April 21, 2011
Address
7405 Westfield Blvd, Indianapolis, IN, 46240

Description

Picasso/Picasso Lite lasers Intended use: Dentistry.

Reason

Product is noncompliant with the Federal laser product performance standard in certain requirements. The ON/OFF key on some of the Picasso/Picasso Lite lasers could be removed with force while the key was in the ON position.

Action

Letters, dated April 30, 2010, addressed to Dental Professionals were sent identifying the affected product and reason for the recall. The letter asked customers to check their units and, if applicable, AMD Lasers would replace the key assembly on their lasers. The service would be provided free of charge including round trip shipping charges to the corporate office. Customers are to call to schedule the service and obtain a RMA number and shipping instructions. Questions should be directed to 1-866-999-2635 or [email protected].

Distribution

Worldwide Distribution

Quantity

3,000 units